44
drying function, dry it after cleaning and disinfection.
Methods:
1.
Spread a clean white paper (white cloth) on the flat table, orient the
product vertically on the white paper (white cloth), and then dry the product
with filtered dry compressed air (maximum pressure 3 bar). No liquid oozing
onto the white paper (white cloth), indicates the drying of the product is
complete.
2.
It can also be dried directly by placing it in a medical drying cabinet
(or oven). The recommended drying temperature is 80℃~120℃ and the time is
15~40 minutes.
Notes:
a)
The drying of the product must be performed in a clean place.
b)
The drying temperature should not exceed 138 °C.
c)
The equipment used should be inspected and maintained regularly.
6.4.4
Inspection and maintenance
6.4.4.1 Inspection
In this chapter, we only check the appearance of the product.
1.
Check the product. If any stain or debris is still visible on the product
after cleaning/disinfection, the entire cleaning/disinfection process must be
repeated.
2.
Check the product. If it is obviously damaged, smashed, detached,
corroded or bent, it must be scrapped and should not be used.
3.
Check the product. If the accessories are found to be damaged, they
must be replaced before use. Replaced new accessories must be cleaned,
disinfected and dried before use.
4.
If the service time (number of times) of the product reaches the
specified service life (number of times), please replace it in time.
6.4.6.2Maintenance
Oil lubrication of sterilized and dried products.
The nozzle of the cleaning lubricant is aligned with the air intake hole at the
end of the contra-angle to inject oil for 1-2 seconds.
Содержание Ai-Motor MotoPex
Страница 1: ...Please read this User Manual carefully before use for operating instructions care and maintenance ...
Страница 4: ...2 ...
Страница 5: ...3 ...
Страница 42: ...40 the contra angle ...
Страница 45: ...43 6 4 4 Drying If the cleaning and disinfection process does not have an automatic ...
Страница 56: ...ZMN SM 034 V1 9 20210430 ...