5
1.7.10
Do not activate the push button on the contra-angle head
during operation or slow down This leads to damage of the head or
instrument separation.
1.7.11
Do not remove the contra-angle head when the handpiece is
rotating. If removed may result in the damage of the gear inside the motor
handpiece
1.7.12
Please confirm whether the file is well inserted and locked
before starting the motor handpiece.
1.7.13
Please set the torque and speed as per the recommended
specifications of file manufacturer.
1.7.14
Error in replacing lithium batteries can lead to unacceptable
risks, so use the original lithium battery and replace the lithium battery
according to the instructions.
1.7.15
Do not place the equipment in a difficult position to avoid difficulty
in disconnection after use,
1.7.16
Please remove the battery if the Endo motor is not likely to be
used for some time.
1.7.17
Wireless charging will generate heat, and the surface temperature
of the charging base and motor handpiece will rise. It is
recommended that the time of contact between the motor handpiece
and charging base during wireless charging should not exceed 10
seconds (only for Ai- Motor).
1.8
Device safety classification
1.8.1
Type of operation mode: Continuous operating device
1.8.2
Type of protection against electric shock: Class II equipment
with internal power supply
1.8.3
Degree of protection against electric shock: B type applied part
1.8.4
Degree of protection against harmful ingress of water: Ordinary
equipment (IPX0)
1.8.5
Degree of safety application in the presence of a flammable
anesthetic mixture with air, oxygen, or nitrous oxide: Equipment cannot
be used in the presence of a flammable anesthetic mixture with air,
oxygen, or nitrous oxide.
1.8.6
Applied part: contra angle, lip hook, file clip, touch probe.
1.8.7
The contact duration of applied part: 1 to 10 minutes.
1.8.8
The temperature of the surface of applied part may reach till
46.6℃.
Содержание Ai-Motor MotoPex
Страница 1: ...Please read this User Manual carefully before use for operating instructions care and maintenance ...
Страница 4: ...2 ...
Страница 5: ...3 ...
Страница 42: ...40 the contra angle ...
Страница 45: ...43 6 4 4 Drying If the cleaning and disinfection process does not have an automatic ...
Страница 56: ...ZMN SM 034 V1 9 20210430 ...