10
2.2.3
Canal Measurement Display
This appears when a file is inside the canal and the lip hook is in
contact with the patient’s lip. Bars on the display show the location of the
file tip in the root canal. In the EAL Mode, If the length is less than 1.0,
the display will be enlarged.
The numbers displayed such as 1.0, 2.0, 3.0 and digital numbers 00-16
do not represent the actual length from the apical foramen. It indicates the
file progression towards the apex. The digital numbers -1 and -2 indicate
that the file extended beyond the apical foramen. The digital number “00”
indicates that the file has reached the apical foramen. Subtract 0.5-1mm
from the measured file length as the working length. These numbers are
used to estimate the canal’s working length.
2.3
Instructions for contra angle
2.3.1
The contra angle adapts precision gear transmission, and the
transmission ratio is 6:1
2.3.2
Before the first use and after the treatment, please clean and
disinfect the contra-angle head with the disinfectant of neutral PH
value. After disinfection, lubricate it with recommended cleaning oil.
Finally, sterilize it under high temperature and high pressure (134℃,
2.0bar~2.3bar (0.20MPa~0.23MPa)). The contra angle should be used
only with this device. The contra angle will be damaged if there is a
mismatch with the device.
Содержание Ai-Motor MotoPex
Страница 1: ...Please read this User Manual carefully before use for operating instructions care and maintenance ...
Страница 4: ...2 ...
Страница 5: ...3 ...
Страница 42: ...40 the contra angle ...
Страница 45: ...43 6 4 4 Drying If the cleaning and disinfection process does not have an automatic ...
Страница 56: ...ZMN SM 034 V1 9 20210430 ...