Reverse osmosis system
GENO-OSMO-X
Order no.
134 750 927-inter Edited by: phar/mrie G:\BA-750927-inter_134_GENO-OSMO_X.docx
47
Parameters
Factory set-
ting
Setting
range
Comment
Switch-on
pressure
(only in case of sys-
tem output Online)
[bar]
3.0
1.0 ... 5.0
In automatic operation, the
system starts the permeate
production if the switch-on
pressure is undershot.
Switch-off
pressure
(only in case of sys-
tem output Online)
[bar]
4.0
1.0 ... 4.0
In automatic operation, the
system terminates the perme-
ate production if the switch-off
pressure is exceeded.
Note: The systems are de-
signed for 4.0 bar.
Automatic
restart
No
No/Yes
What to do after a power
failure
No
= after return of power, the
system signals "Error power
failure“; automatic resp. man-
ual operation remains stored
but is blanked.
YES
= after return of power,
the system continues running
in the operating mode set pre-
viously and does not indicate
an error.
Forced operation
Operation
Operation/
flushing
Behaviour in case of a longer
period without permeate pro-
duction (also refer to the cus-
tomer service level):
Operation
= permeate pro-
duction, switch-off command
tank full is ignored (tank with
overflow required)
Flushing
= Factory setting in
case of system output Online:
Flushing to the drain, high-
pressure pump is not running.
Conductivity
monitoring
RO1CQ1
Signal
Signal/
error
Monitoring of permeate con-
ductivity:
If the programmed conductivity
limit value is exceeded for the
set delay time, the following
options can be programmed:
Signal
: The system continues
running although the limit
value was exceeded.
Error
: The system switches
off.
Conductivity
limit value
RO1CQ1
Cell constant 0.1
[µS/cm]
30
0 ... 99
Conductivity limit
value
RO1CQ1
Cell constant 1.0
[µS/cm]
30
0 ... 999
Delay conductivity er-
ror/signal
RO1CQ1
[min.]
30
0 ... 999