Reverse osmosis system
GENO-OSMO-X
Order no.
134 750 927-inter Edited by: phar/mrie G:\BA-750927-inter_134_GENO-OSMO_X.docx
41
Note:
Prior to removing the SD card, make sure that the data
logging was terminated with "Nein/No"!
Note:
The SD card used must be formatted FAT32. We recom-
mend proper formatting instead of quick formatting.
Note:
The SD card socket is integrated in the operating element
of the GENO-tronic - to insert the card, the control unit of the
GENO-OSMO-X must be opened.
Parameters
Factory
setting
Setting
range
Comment
Operating language
German
German
English
French
Dutch
Italian
Russian
Spanish
System designation
18 digits, alphanumerical,
text will appear in the upper
left corner of the Home
screen
Date, time
Incl. automatic switch-over
from daylight saving time to
standard time
Data logging
Start
Start/
Stop
All measured values of all
subsystems (Info level) are
logged on the SD card.
Interval
[min]
1 ... 999
Indicates the frequency of
how often the measured val-
ues are stored on the SD
card
Load parameters
Before doing so, load the pa-
rameter record stored on the
SD card
Attention: Do not execute
during permeate production!
Save parameters
Save current parameter
record on the SD card
Screensaver
[min]
1
0…99
Lock screensaver
[sec]
30
10…99
Note:
The protective foil over the touch screen may only be
wiped with a damp cloth and a mild soapsuds solution at the
most! Never use cleaning agents for plastics or stronger clean-
ing agents! Otherwise there is a risk that the foil will assume a
milky colour and therefore be difficult to read!