Information for the doctor and qualified medical staff 83
Auto-ID
This displays the ID number that is automatically assigned every time
the
Admit new patient
function is executed.
Date, time
This displays the date and time of the internal clock which can be set
in the
System
menu.
9.2.3
Info\ Measurements: SpO
2
Info\ Measurements:
SpO
2
displays various average values
for
SpO
2
calculated since the
monitor was switched on.
These values are lost when the
monitor is switched off.
Fig. 55 Info\ Measurements: SpO
2
SpO
2
:
Average
is calculated over the
Averaging interval
set in the
SpO
2
menu. The
Current deviation
shows how much the current
SpO
2
deviates from the
Average
in percent. This deviation is used for
reporting deviation alarms when
SpO
2
Alarms
in the
SpO
2
menu has
been set to
Limits & trends
.