102 Algorithms and measuring principles
10 Algorithms and measuring principles
Knowledge of the following calculation bases is essential if VitaGuard®
is to be properly configured.
10.1 Alarm condition and report delays
As prescribed in the standard IEC 60601-1-8 “General requirements
for safety – Collateral standard: General requirements, tests and
guidance for alarm systems in medical electrical equipment and
medical electrical systems”, this section allows the doctor to become
familiar with the set and inherent delays for the correct configura-
tion of the alarm limits and monitoring parameters.
The alarm
condition
delay is the time from the occurrence of a
triggering event on the patient or in the monitor to the decision
by the alarm system to confirm an alarm condition.
The alarm
report
delay is the time between when an alarm condi-
tion is detected and when it is reported.
The alarm
condition
delay and the alarm
report
delay are added to
yield the ALARM SYSTEM delay.
The algorithms listed here are based on worst credible cases, i.e. the
calculations always return the maximum possible delays.
An alarm
condition
delay as defined in the standard does not occur
when oxygen saturation is monitored. Please observe though the
information on alarm
report
delays below.
10.2 Alarm report delays
The alarm report delays for bradycardia, tachycardia, hypoxia, and
hyperoxia can be set within certain limits.