Alarm and monitor settings 77
group alarm limits to the current patient. These settings can be
changed only when
Settings protection
has been set to
Limited
in
the
System
menu.
Different pulse rate alarm limits can be set as the default values for
each age group:
Default
0 to 2 years
2 to 6 years
>
6 years
Lower pulse rate limit [/min]
80
60
55
Upper pulse rate limit [/min]
220
150
140
The pulse rate’s
Lower limit
can
be set from 30 to 180 beats and
the pulse rate’s
Upper limit
from
100 to 255, each in steps of five
beats per minute.
Factory settings are shown in
bold type.
Fig. 52 Pulse rate menu for viewing and setting alarm limits
Lower limit
(pulse rate) ..... 30, 35 ...
80
... 175, 180/min
Lower limit for the pulse rate; an alarm
is reported when the pulse rate falls be-
low this limit for longer than the set
Bradycardia delay
Upper limit
(pulse rate) ...... 100, 105 ...
220
... 250, 255/min
Upper limit for the pulse rate; an alarm
is triggered when the pulse rate ex-
ceeds this limit for longer than the set
Tachycardia delay