66 Alarms, displays, and views during monitoring
Message
Meaning
Cause or elimination
No.
SpO
2
:
Unrecog-
nized
sensor!!
The SpO
2
module reports
that an unrecognized
sensor is connected.
-
Replace the SpO
2
sensor
(use only the sensors from Masimo Inc.).
29
7.12 Table of information messages
Message Cause
Meaning
No.
Calculating
pulse rate
The current pulse rate
cannot be displayed
while it is being
calculated.
The current pulse rate is displayed
after it has been calculated.
37
Internal battery
too low
The internal battery for
alarms during a power
failure is depleted.
-
The monitor is defect i. e. the
internal battery needs to be replaced
by a technician.
39
SpO
2
: Low
perfusion
The SpO
2
module
reports that the blood
flow is too weak.
-
Either use a different attachment
site, or set
Sensitivity
to
maximum
in the SpO
2
menu.
35
SpO
2
: Low signal
IQ
The SpO
2
module
reports that the signal
quality is low.
-
Use a different attachment site, or
check for the presence of light or
electromagnetic interference
sources in the vicinity.
-
Whenever possible, prevent vigorous
movements by the patient.
36
Status: ok
No messages
40