Alarms, displays, and views during monitoring 65
Message
Meaning
Cause or elimination
No.
No cables
connected!!
The monitor discovers
that both patient cables
are not connected.
-
Connect the patient cable.
21
No power
adapter !!
The power adapter has
been disconnected.
-
Reconnect the external power adapter
or press the <Esc> key.
17
Recharge
battery!!
The battery voltage is too
low: the monitor can no
longer operate reliably
-
Operate the monitor with the external
power adapter to recharge the block
battery, or insert non-rechargeable
batteries.
32
Replace
batteries!!
The battery voltage is too
low: the monitor can no
longer operate reliably.
-
Insert new batteries or a new block
battery or operate the monitor with
the external power adapter.
33
SpO
2
: Check
cable!!
The SpO
2
module reports
that the SpO
2
cable is not
connected.
-
Connect the SpO
2
cable.
-
Replace the SpO
2
cable, if this message
persists.
26
SpO
2
:
Defective
sensor!!
The SpO
2
module reports
that the SpO
2
sensor is
defect.
-
Replace the SpO
2
sensor.
27
SpO
2
:
Hardware
fault!!
The SpO
2
module is not
supplying data.
-
Switch off the monitor, wait for thirty
seconds, and switch it back on: if this
message persists, the monitor is defect.
15
SpO
2
:
Interfer-
ence!!
The SpO
2
module detects
electromagnetic interfer-
ence.
-
Locate any interference sources in the
direct vicinity and, if necessary, remove
them.
30
SpO
2
: Pulse
search!!
When switched on, at
first, the SpO
2
module
reports that it is search-
ing for the pulse.
-
When this message is displayed during
monitoring, check whether the sensors
are secured and positioned properly.
34
SpO
2
:
Sensor off!!
The SpO
2
sensor is defect
or not connected.
-
Check whether the SpO
2
sensor is
correctly connected to the cable; if
necessary replace the sensor.
28
SpO
2
: Too
much
light!!
The SpO
2
module reports
that there is too much
light.
-
Protect the SpO
2
sensor from light
sources, e.g. by covering it.
31