Alarm and monitor settings 75
no longer monitored. When the
Sensitivity
in the SpO
2
menu is set to
Maximum
, the cut-off limit is 0.02%; when set to
Standard,
this limit
ranges from 0.5 to 0.02%, depending on the signal quality.
The principal of operation for this calculation can be found in the
section “Measuring principle for the SpO
2
monitor” on page 103.
8.6.2
SpO
2
menu – alarm settings (
Settings protection
Limited
)
The
SpO
2
menu lets you view and
change the current SpO
2
alarm
settings.
Changes are permitted only
when
Settings protection
is set
to
Limited
in the
System
menu.
Factory settings are shown in
bold type.
Fig. 50 SpO
2
menu for viewing and setting alarm limits
Lower limit
............................ 50, 51 ...
88
... 99, 100 %
Lower alarm limit for the measured ar-
terial oxygen saturation; an alarm is re-
ported when the measured value falls
below this limit for longer than the set
Hypoxia alarm delay
Upper limit
............................. 50, 51 ... 88 ... 99,
100
%
Upper alarm limit for the measured ar-
terial oxygen saturation; an alarm is re-
ported when the measured value ex-
ceeds this limit for longer than the set
Hyperoxia alarm delay
Содержание VitaGuard VG310
Страница 1: ...VitaGuard VG310 Pulse oximeter Operating instructions...
Страница 2: ......
Страница 4: ......
Страница 10: ...Table of contents...
Страница 108: ...108 Evaluating stored data on a PC Fig 79 VitaWin register Events in graph form...