110 Specifications
Characteristics of
acoustic alarm signals ......... acoustic signals for higher-priority
alarms consist of two acoustic se-
quences of five tones each:
Pulse duration
-
155 ms
±
5 ms
Rise and fall time
-
17 ms
±
3 ms
Time between start of pulse
1 and start of pulse 2
T
1
215
ms
±
20 ms
Time between start of pulse
2 and start of pulse 3
T
1
215
ms
±
20 ms
Time between start of pulse
3 and start of pulse 4
T
2
440
ms
±
20 ms
Time between start of pulse
4 and start of pulse 5
T
1
215
ms
±
20 ms
Time between start of burst
1 and start of burst 2
T
3
2
s
±
0.1 s
Repeat time for the whole
sequence
- 10
s
±
0.2 s
acoustic signals for medium-priority
alarms consist of three-pulse bursts:
Pulse duration
185 ms
±
5 ms
Rise and fall time
17 ms
±
3 ms
Time between start of pulse
1 and start of pulse 2
375 ms
±
20 ms
Time between start of pulse
2 and start of pulse 3
375 ms
±
20 ms
Repeat time for the whole
sequence
5.2 s
±
0.2 s