Alarms, displays, and views during monitoring 61
the period set under
Hypoxia alarm delay
or exceeds the SpO
2
Upper
limit
for longer than the period set under
Hyperoxia alarm delay,
an
acoustic alarm signal is emitted and the corresponding message is
displayed. The affected alarm limit and the alarm LED flash.
Go immediately to the patient when an alarm occurs and check the
patient’s condition.
When the SpO
2
value returns within the permitted range the alarm is
ended automatically. In this case, the affected alarm limit and the
alarm LED continue to flash until the <Esc> key is pressed to indicate
that an alarm has occurred.
The SpO
2
Upper limit
is deactivated when it is set to 100% (factory
setting). We recommend setting an upper limit when the patient is
undergoing oxygen therapy.
7.9.2 Technical
SpO
2
alarms
The section “Table of technical alarm messages” on page 64 can be
consulted for the technical alarm signals and the recommended
troubleshooting procedures.
The SpO
2
monitor displays technical alarms with the corresponding
messages. Until the problem has been eliminated, the SpO
2
value and
the pulse rate are replaced by a question mark symbol. Perfusion and
signal IQ are set to zero.
7.10 Pulse
rate
alarms
After the monitor is switched on, it may take up to twenty seconds
before the first values are displayed.
The currently set alarm limits are always displayed.