72 Alarm and monitor settings
8.5.5
System\ Alarm tone pitch
You can set the pitch of the acoustic alarm signals to
Low
,
Medium
,
or
High
so that they can be heard over the expected background
noise. The DIN settings (DIN) match the alarm tone characteristics as
described from page 58 on in the section “Differentiating physiologi-
cal and technical alarm signals”. As
an alternative, you can set the alarm
tone characteristic (gtm) as familiar
from other getemed devices. The
factory setting is
Medium
.
Fig. 47 System\ submenu “Alarm tone pitch”
8.5.6
System\ RS232 format
This submenu lets you assign the
format for online data output from
the AUX serial port.
Fig. 48 System\ RS232 format
8.5.7
System\ Settings protection
On,
Limited, Off
The codes that protect the alarm defaults from unauthorized
changes must be given by the doctor to those persons only whom
the doctor judges to be adequately informed about monitoring and
their responsibility for the patient. The doctor should point out that
the code must be treated as confidential, that settings should be
changed at the doctor’s request only, and that all changes must be
confirmed by the doctor.
VitaGuard® provides the following three settings for
Settings pro-
tection.