76 Alarm and monitor settings
8.7
Pulse rate
display and menu
The
Z
key takes you from
View
1, or 2 to the pulse rate display. From
here you can open the menu with the
U
or
V
key. When a row is
highlighted, press <Enter> to change the corresponding value.
8.7.1
Pulse rate
display
The top half of the display
presents:
1 the status line
2 the current values with
the set alarm limits
3 the current three-minute
trends that update the last
value every two seconds
Fig. 51 Pulse rate display
The pulse rate trend display presents the pulse rate over the last three
minutes. This display varies with the set age group:
0
to
2
years
2
to
6
years
>
6
years
Pulse rate
trend display
[per min]
between 230 and 50 between 180 and 50 between 150 and 45
The bottom half of the display presents the plethysmogramme. A small
vertical bar indicates every detected pulse.
8.7.2
Pulse rate
menu – alarm settings (
Settings
protection Limited
)
The
Pulse rate
menu lets you view and, if necessary, change the
current pulse rate settings. You may have to adapt the default age
1
2
3