GLX-RSS-2-300 WL User Manual
27
GLX-RSS-2-300 WL Non-Contact Flow Meter
Water level report
$LVL,L1,L2,T1,L3,L4,S1,S2*CSUM<CR><LF>
$LVL:
The keyword sent in the beginning of each level report.
L1:
Current distance from sensor to water, in defined units, 0 being sensor
reference plane. If reporting -4 then there is no detected level.
L2:
Average distance from sensor to water, in defined units, 0 being sensor
reference plane. If reporting -4 then there is no detected level.
T1:
Internal device temperature (in °C).
L3:
Current relative detected level, in defined units, 0 being sensor reference
plane.
L4:
Average relative detected level, in defined units, 0 being sensor reference
plane.
S1:
Measurement SNR (in dBm).
S2:
The standard deviation of water level measurements. The number of
samples used to calculate the standard deviation is equal to the configured
filter length.
CSUM:
The check sum of the characters in the report from $ to * excluding these
characters.
Level sensor tilt angle report
$LVLANG,A1,A2 *CSUM<CR><LF>
$LVLANG:
The keyword sent in the beginning of each level sensor tilt angle report.
A1:
The tilt angle of the instrument along X axis. For proper operation of the
instrument, ensure that this angle is zero, or as close to zero as possible
(between -1 and +1 degrees).
A2:
The tilt angle of the instrument along Y axis. For proper operation of the
instrument ensure that this angle is zero, or as close to zero as possible
(betwee n -1 and +1 degrees).
CSUM:
The check sum of the characters in the report from $ to * excluding these
characters.