GLX-RSS-2-300 WL User Manual
15
GLX-RSS-2-300 WL Non-Contact Flow Meter
Warm Up Time
Warm up time
- The time after sensor power-up, during which all measurements are ignored. This time
is used to settle auto-gain parameters, Kalman filter values, averaging filter, and all other operational
parameters. It is recommended to set this value to a minimum of 5 seconds. In extreme cases where
a quick response after unit power-up is required, 3 seconds can be used, with a possibility of losing
measurement accuracy.
4 – 20 mA output
4 – 20
mA output
- This parameter is used to select the value that will correspond to the 4 – 20 mA
output. When
Velocity
is selected, the output current will be proportional to the measured velocity. When
Level
is selected, the output current will be proportional to the measured level and when
Discharge
is
selected, the output current will be proportional to the calculated discharge. When
None
is selected, the
4 – 20 mA output will be disabled.
4 – 20 mA min. and 4 – 20 mA max.
4 – 20 mA min.
– To configure the 4 – 20 mA output range, the minimum measured value which
will correspond to 4 mA analog output needs to be set. The value is set in the currently configured
measurement unit. Example: if values measured by the instrument are expected to be within the range
of 700 mm/s up to 1500 mm/s, it is recommended to configure the minimum value to slightly below 700
mm/s (for example 500 mm/s). Alternatively, if the resolution is not critical, then the minimum value for
4 – 20 mA output can be left to the instrument minimum of 0 mm/s.
4 – 20 mA max.
– To configure the 4 – 20 mA output range, the maximum measured value which
will correspond to 20 mA analog output needs to be set. The value is set in the currently configured
measurement unit. Example: if values measured by the instrument are expected to be within the range
of 700 mm/s up to 1500 mm/s, it is recommended to configure the maximum value to slightly above
1500 mm/s (for example 2000 mm/s). Alternatively, if the resolution is not critical, then the maximum
value for 4 – 20 mA output can be left to the instrument maximum of 15000 mm/s.
WARM UP TIME
POWER ON
Dead time
Output value 4-20 mA
20 mA
4 mA
Min. Value
Max. Value