GLX-RSS-2-300 WL User Manual
3
GLX-RSS-2-300 WL Non-Contact Flow Meter
3
Cable Pin-Out
The flow meter is supplied with open end cable consisting of 12 wires coded with colours. Table 2
gives a detailed description for each wire.
It is possible to mount optional connectors on the cable end per user request. In this case, the
connector specification and connection diagram must be specified when ordering the sensor.
Table 2. Cable Pin-Out
Pin No.
Wire Color
Pin Name
Pin Description
1
White
GND
This pin should be connected to the ground (neg
-
ative) pole of the power supply.
2
Brown
+Vin
The power supply for the Radar Speed Sensor is
provided on this pin. The Radar Speed Sensor pow
-
er supply voltage must be in the range 9 VDC to 27
VDC, and the power supply must be able to provide
at last 0.65W
3
Green
RS232 – TxD
RS-232 data transmit signal.
4
Yellow
RS232 – RxD
RS-232 data receive signal.
5
Grey
GND
Signal ground.
6
Pink
CAN – H
CAN2.0B high signal. (optional)
7
Blue
CAN – L
CAN2.0B low signal. (optional)
8
Red
SDI12 DATA
SDI12 data line
9
Orange
RS485 – D-
RS-485 data transmitter/receiver low signal.
10
Dark Red
RS485 – D+
RS-485 data transmitter/receiver high signal.
11
Black
Alarm SW or 4–20 mA
Secondary (Optional)
Alarm 1 - open collector switch signal max.
60mA
12
Purple
4 – 20 mA
Sink for 4 – 20 mA analog interface. Connect
sensing device as pull-up to sink the current