GLX-RSS-2-300 WL User Manual
16
GLX-RSS-2-300 WL Non-Contact Flow Meter
Level 4 – 20 mA min. and 4 – 20 mA max.
Level 4 – 20 mA min.
–To configure the 4 – 20 mA output range, the minimum measured value which
will correspond to 4 mA analog output needs to be set. The value is set in the currently configured
measurement unit. Example: if values measured by the instrument are expected to be within the range
of 700 mm up to 5000 mm, it is recommended to configure the minimum value to slightly below 700
mm (for example 500 mm). Alternatively, if the resolution is not critical, then the minimum value for
4 – 20 mA output can be left to the instrument minimum of 0 mm.
Level 4 – 20 mA max.
– To configure the 4 – 20 mA output range, the maximum measured value which
will correspond to 20 mA analog output needs to be set. The value is set in the currently configured
measurement unit. Example: if values measured by the instrument are expected to be within the range
of 700 mm up to 5000 mm, it is recommended to configure the maximum value to slightly above 5000
mm/s (for example 6000 mm). Alternatively, if the resolution is not critical, then the maximum value for
4 – 20 mA output can be left to the instrument maximum.
Output value 4-20 mA
20 mA
4 mA
Min. Value
Max. Value