49
GSM19 Series Magnetometers - Version 7
GEM Systems, Inc. Advanced Magnetometers
For more technical information, visit www.gemsys.ca
To
Stop Reading
or
Exit
the
Reading Menu
,
press A
. The following screen is displayed:
SCREEN 36
A-position B-enter text
4-graph vertical scale
C-tune
5-display-mode
E-EOL F-ok 0-noise
1-info
The following menu gives you access to some of the parameters in the set up menus.
•
A-position
- Will take you to the position coordinates setup screen. NOTE that from this menu
you can only change the coordinates but you cannot change the system (see section 5.3.2).
•
B-enter text
- Allows you to type messages or comments which can later be recovered with
SEND.
•
4-graph
- Is active only if graph mode is selected. It allows to change the graph's scale and
offset (see section 5.3.8)
•
C-tune
- Leads you to the tuning setup screens (see section 5.3.6). NOTE that the tune initialize
process is not repeated when returning to survey (even when set to YES).
•
5-display-mode
- Guides you through the selection of Display Mode (see section 5.3.8).
•
E-EOL
- Performs
End Of Line
function - line number is incremented, direction is reversed
(see section 5.4.4).
•
F-ok
- Takes you back to survey.
•
0-noise
- Will display noise from sensor - value of 100 or less is normal (see section 5.4.3).
•
1-info
- Is very useful in providing information about last reading like s/n (signal/noise ratio)
and measurement time in ms. It also shows the number of readings which can be taken before
the memory is full (see section 5.4.5).