143
GSM19 Series Magnetometers - Version 7
GEM Systems, Inc. Advanced Magnetometers
For more technical information, visit www.gemsys.ca
If the start time has already passed, the unit starts immediately. Otherwise it displays:
SCREEN G9
050916
waiting for program
07:00 to 19:00
•
Where 050916 is the current time in hhmmss.
•
When the start time is reached, the survey begins.
NOTE:
The start time indicates the beginning of the survey and NOT the time of the first
reading. Depending on the selected cycling rate the time of the first reading of the
file is delayed. Furthermore, to preserve the battery the GPS module is turned off
while waiting for a program start. If GPS is enabled, the unit turns on GPS and
waits for GPS lock and UTC synchronization. This imposes a further delay on the
time required for the first reading.
At the stop time, the unit stops cycling, turns off GPS if enabled, closes the current file and waits
for the start time the next day.
The daily survey loop can only be stopped by pressing
1C
either while taking readings or while
waiting for the start time.
I.4 USING THE FLEXIBLE SCHEDULING FEATURE
The flexible scheduling feature lets you store up to 30 schedules each with date and time for the
start and the date and time to stop the surveys.
SCREEN G10
.base start
A-immediate
B-daily
C-programmable D-RS-232
Select
C-programmable.
A screen similar to the following is displayed.