Chapter 3 Installation
3-13
3-5-3
Transducer Connection
1.) Connect a transducer to the upper transducer receptancle as follows:
A.) Ensure that the transducer twist lock lever points towards the 9 o’clock position.
B.) Insert the transducer connector on the receptacle guide pin until it touches the receptacle mating
surface.
C.) Twist the transducer lock lever vertically to lock it in place. Twist the lever to the horizontal position
to disconnect the transducer.
NOTE:
Please ensure that the probe latch is in unlocked position before you connect the probe to the system.
NOTE:
It is not necessary to turn OFF power to connect or disconnect a probe.
Figure 3-8 Connect the probe
Fig (A) Connecting the Probe
Fig (B) Locking the Probe Lever