Section 8-2 - Disassembly/Re-assembly of VIVID P3
Alpha Numeric Keyboard with cables (FRU P/N :5340669)
3) Remove Trackball assembly with Trackball Interface cable. Refer
Figure 8-16 Optical Trackball Assembly
8-2-10-6
Mounting procedure
1.) New parts in the reverse order of removal.
8-2-10-7
Functional Checkout Procedure
See
Section
Functional Test
Debrief Script
Service Manual Direction 5344303-100, Section 8-2-9. Equipment passes all required
tests and is ready for use.