GE H
EALTHCARE
PROPRIETARY TO GE
D
IRECTION
55344303-100, R
EVISION
3
V
IVID
P3 S
ERVICE
M
ANUAL
Chapter 5 Components and Functions (Theory)
5-21
Section 5-3
Power Diagrams
5-3-1
Overview
The AC Power assy’s main tasks are to isolate and output to the AC/DC unit which is inside the system
console. The input of AC power pack will be the AC outlet, the range is AC 100v Min & 240v Max
configuration support is available in power pack.
Figure 5-16 AC Power Distribution Block Diagram (Vivid P3)
The mains cord has plugs in one side end (On Wall). A male plug (of Power Cord) connects to the mains
outlet on site.
From the Main Circuit Breaker, the AC power is routed via an Inrush Current Limiter to a internal outlet
connector for the Main power supply. The power ON OFF is actually controlled by the MST PWA. This
signal is routed from the Keyboard ON OFF button to the MST board and then to the PDB board.
The MST board has logic that controls the state of this signal. The output from this logic is connected
to the PDB board that either switches on or switches off the Power supply.