Chapter 4 Functional Checks
4-29
4-3-12
ECG Checks
4-3-12-1
Introduction
The ECG capability on this unit, is intended as use as a trigger for measurements, but can also be viewed on the
Screen.
4-3-12-2
ECG Check
Table 4-13 ECG Check
Step
Task
Expected Result(s)
1
Connect the ECG harness to the connector on the
front of the system
The unit displays a straight curve along the bottom
edge of the image sector on the screen.
2
Connect the three leads to a ECG simulator,
or
Fasten the three ECG Pads to your body and connect
the three leads to respective ECG Pad
When connecting, the signal on the screen will be
noisy. When the connection is completed, a typical
clean ECG signal is displayed.