GE H
EALTHCARE
PROPRIETARY TO GE
D
IRECTION
5344303-100, R
EVISION
3
VIVID P3 S
ERVICE
M
ANUAL
8-4
Section 8-2 - Disassembly/Re-assembly of VIVID P3
8-2-4
LCD Assy (FRU P/N: 5315112)
Purpose: This is a description on how to remove and replace the LCD Assy.
8-2-4-1
FRU BOM
This FRU Kit P/N:5315112 consists of following parts.
8-2-4-2
Tools
•
Common phillips screwdrivers
•
Allen/Unbraco wrench
8-2-4-3
Needed Manpower
•
1 person, 11 m travel
8-2-4-4
Preparations
•
Shut Down the System Switch of the circuit breaker at the bottom rear side of the system.
8-2-4-5
Removal procedure
1) Shut down the Vivid P3 system.
2) LCD monitor rearcover of Vivid P3 system.as shown below
Figure 8-1 LCD rear cover
3) Bend down the LCD monitor by 90 degree.
4) Bend the LCD monitor as shown below
PART NO
DESCRIPTION
5245083
LCDMON II
5314204
Fasteners-M5 X 16 Socket Head Screw
5314209
Monitor Label Vivid P3
LCD Rear cover
LCD Arm cover
LCD power
cable cover
Содержание Vivid P3
Страница 2: ...This page was intentionally left blank...
Страница 9: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5344303 100 REVISION 3 VIVID P3 SERVICE MANUAL vii...
Страница 292: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5344303 100 REVISION 3 VIVID P3 SERVICE MANUAL 9 8 Section 9 6 Mechanical Assy...