Section 4-3 - General Procedure
4-3-17-2
Backup System Presets and Configurations
NOTE:
Always backup any preset configurations before a software reload. This ensures that if the presets need
to be reloaded, after the software update, they will be the same ones the customer was using prior to
service.
4-3-17-3
Patient Healthcare Information (PHI)
Patient Healthcare Information (PHI) is the patient data stored in the USB pendrive/USB HDD inserted
in the VIVID P3 system. Ensure that the PHI is erased from the USB pendrive/USB HDD, or the SATA
HDD is removed from the VIVID P3 system before shipping (repairs/replacement).
In case, if any patient information still resides on the VIVID P3 system, GE contact the customer to
immediately collect the data. However, GE keeps the patient information in a secured environment for
a period of one month. The data is deleted permanently after one month.
It is strictly prohibited to send the PHI outside the confidential patient-doctor environment without
encrypting the data. If PHI is sent to GE employees for troubleshooting, the PHI must be encrypted or
protected from unauthorized access and the files must be compressed using PKZIP with a password.
The PHI must be sent to GE employees only through the GE intranet. For VIVID P3, the GE intranet
address is http://libraries.ge.com/foldersIndex.do?entity_id=12065698101&sid=101&sf=1
NOTE:
PHI (Patient HealthCare Information) data should not be sent to the GE Employees.
It is not allowed to send the data outside of confidential Patient-Doctor environment without encryption
of PHI. If PHI needs to be sent to GE employees for troubleshooting purposes, all PHI should be
encrypted or protected from unauthorized acess. Do not email the data which includes ‘PHI’. It can only
be sent via GE Internet (ONLY FTP).
1.) Insert a formatted DVD-R into the drive.
2.) On the Keyboard, press
CONFIG
.
3.) On the LCD display, press
SYSTEM
.
4.) On the LCD display, select
BACKUP/RESTORE
.
NOTE:
If you are not logged in as GE Service or with administrator privileges, the Operator Login
window is displayed. Log on with administrator privileges (By default, the password is
“ulsadm”).
5.) In the Backup list, select Patient Archive, Report Archive and User Defined Configuration.
6.) In the Media field, select DVD-RW (or USB memory device).
7.) Select
BACKUP
.
The system performs the backup. As it proceeds, status information is displayed on the Backup/Restore
screen.