Chapter 8 Replacement Procedures
8-55
Section 8-5 USB Cable Set (FRU P/N: 5315039)
8-5-0-1
Tools
•
Common Phillips screwdrivers
8-5-0-2
FRU BOM
This FRU Kit P/N:5315039 consists of following parts.
8-5-0-3
Needed Manpower
•
1 persons, 15 minutes per cable
8-5-0-4
Preparations
•
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-5-0-5
Removal Procedure
1. Remove the covers and parts as applicable. Refer to the applicable section of this document.
2. Unplug the USB cable from both the ends and remove the cable from the machine.
8-5-0-6
Functional Checkout Procedure
Part Number
Part Description
5262527
MST To Keyboard- 4FTAA
5267540
RPI to Printer- 3FT AB2
5262590
Docking Board to Top panel 4FT AA
5269176
RPI to ECG- 3FT AA2
5270287
MST to RP1- 1FT AB
See
Section
Functional Test
Debrief Script
Service Manual Direction 5344303-100, Section 8-5. Equipment passes all required
tests and is ready for use.