Section 8-12 - Power Cable Set (FRU P/N: 5315036)
8-12-1
CWD Assy (FRU No:5315026 )
Purpose: This is a description on how to remove and replace the CWD Assy.
8-12-1-1
Tools
•
Common phillips screwdrivers
•
Allen/Unbraco wrench
8-12-1-2
Needed Manpower
•
1 person, 30 m travel
8-12-1-3
Preparations
•
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
NOTE:
If the CWD Board is Defective/Not Present in the image mode “Mode not available on this
probe“ message will apear.
Removal Procedure of CWD. Refer to Figure 8-71.
Figure 8-71
1.) Unscrew two screws(1-2).
2.) Disconnect the CWD assy.
8-12-1-4
Removal procedure
Refer to
1.) Unscrew two screws(1-2) to remove the CWD cover. see
2.) Connect the CWD assy to RX64. [See
3.) Press the CWD assy tightenedly, refer to
.
4.) Screw two screws(1-2) to fix the CWD assy on RX64 assy., see
]