GE H
EALTHCARE
PROPRIETARY TO GE
D
IRECTION
5344303-100, R
EVISION
3
VIVID P3 S
ERVICE
M
ANUAL
8-14
Section 8-2 - Disassembly/Re-assembly of VIVID P3
Keyboard Assembly (FRU P/N :5340674)
5) Remove cable from keyboard Assembly of the system. Refer
Figure 8-13 Keyboard cable location
8-2-7-5
Mounting procedure
1.) Install the new parts in the reverse order of removal.
8-2-7-6
Replace Keyboard Assembly
1.) Connect keyboard cable to the new Keyboard Assembly.
2.) Locate the Keyboard top hinge appropriately.
3.) Replace keyboard Assembly by fixing three screws (a,b,c).
4.) Refix Gel holder and Probe holders.
8-2-7-7
Functional Checkout Procedure
See
Section
Functional Test
Debrief Script
Service Manual Direction 5344303-100, Section 8-2-6. Equipment passes all
required tests and is ready for use.