4-16
Section 4-3 - General Procedure
4-3-8
Color Flow Mode Checks
4-3-8-1
Preparations
1.) Connect one of the probes listed in
3-6-5 "Available Probes" on page 3-24
, in Chapter 3 Installation
to the System probe connector.
2.) Turn ON the scanner (if it isn’t turned on already).
Figure 4-12 Controls available in Color Flow Mode
Figure 4-13 CFM Mode Screen Picture Example
Primary Menu
CF Mode
Secondary Menu
Frequency
Angio
Line Density
Variance
Baseline
Packetsize
Scale
Map
Invert
Low Vel Reject
Threshold
Spatial Filter
Flash Suppression
Transparency Map
Focus Position
Frame Average
Power Output
Gain