Section 7-6 - Troubleshooting
Section 7-6
Troubleshooting
7-6-1
Console Troubleshooting Trees
7-6-1-1
System Doesn’t Boot
This is an overall diagram showing a recommended sequence for troubleshooting a no-boot situation.
Figure 7-6 System Doesn’t Boot (cont’d)
Is the power On/Off LED in
Amber color lit?
YES
Check /-10%at i/p
available to the transformer by
conforming whether stand by
led is lit.
NO
Does power On/Off LED is
Green ?
NO
Check main cable between KBD
and PDB board.
Are the LEDs on PDB
module lit?
NO
YES
1) Check cable connections from KBD to PDB
2) If problem persists, replace PDB Module &
the Cable b/w PDB & KBD.
System doesn’t boot!
Replace PDB board.
NO
Go on to the next page.
YES
Press ON/OFF Button
NOTE: Issues Includes
1)System does not Switch ON OR System does not complete
the bootup process.