Probes
Vivid i User’s Manual
271
2378958-100 Rev. 02
Care and Maintenance
This section covers:
• Planned
maintenance
• Probe
inspection
• Probe
cleaning
• Probe
disinfection
Planned maintenance
It is recommended to keep a maintenance log and note all
probe malfunctions. Follow the maintenance schedule below to
ensure optimum operation and safety:
After each use:
•
Inspect the probe
•
Clean the probe
•
If required disinfect the probe
Before each use:
•
Inspect the probe
CAUTION
Improper handling can lead to early probe failure and electric
shock hazards.
DO follow the specific cleaning and disinfection procedures
provided in this chapter and the germicide manufacturers
instructions.
Failure to do so will void probe warranty.
CAUTION
Transesophageal and intraoperative probes require a special
handling. Refer to the user documentation enclosed with these
probes.
Содержание Vivid i
Страница 110: ...Scanning Modes 100 Vivid i User s Manual 2378958 100 Rev 02...
Страница 111: ...Purposely left empty Vivid i User s Manual 101 2378958 100 Rev 02 Chapter 4 Purposely left empty For Future purposes...
Страница 112: ...Purposely left empty 102 Vivid i User s Manual 2378958 100 Rev 02 To Purchase Visit Avobus com or call 1 800 674 3655...
Страница 161: ...Purposely Left Empty Vivid i User s Manual 151 2378958 100 Rev 02 Chapter 7 Purposely Left Empty For Future purposes...
Страница 162: ...Purposely Left Empty 152 Vivid i User s Manual 2378958 100 Rev 02...
Страница 288: ...Probes 278 Vivid i User s Manual 2378958 100 Rev 02...
Страница 350: ...User maintenance 340 Vivid i User s Manual 2378958 100 Rev 02...
Страница 376: ...Safety 366 Vivid i User s Manual 2378958 100 Rev 02 To Purchase Visit Avobus com or call 1 800 674 3655...
Страница 387: ......
Страница 388: ...To Purchase Visit Avobus com or call 1 800 674 3655...