Basic scanning operations
Vivid i User’s Manual
59
2378958-100 Rev. 02
Using cineloop
Selection of a cineloop
1.
Press
FREEZE
.
The left and right markers on the ECG trace are displayed
on either side of the last detected heart cycle.
2.
Press the
2D FREEZE
button to un-freeze the cineloop and
let it run between the default setting of left and right borders.
3.
To modify setting of the left or right borders, press the
CINELOOP
assignable.
The selected heart beat is played back.
4.
Press on the assignable
CYCLE SELECT
to move from heart
beat to heart beat to select the heart cycle of interest.
To jump directly to
the first or to the
last heart beat press
the assignables
FIRST CYCLE
or
LAST CYCLE
.
5.
Press on the assignable
NUM CYCLES
to increase or
decrease the number of heart beats to be played back.
6.
Adjust
LEFT MARKER
and
RIGHT MARKER
assignables to trim
or expand the cineloop boundaries.
Adjustment of cineloop playback
1.
If in freeze mode, press the assignable
2D FREEZE
to start
cineloop replay.
2.
Use the
Trackball
to increase or decrease the speed of the
cineloop playback.
The speed factor is displayed on the right side of the ECG
(see Figure 2-4).
To view a cineloop frame by frame
1.
If not in freeze mode, press the
2D FREEZE
button to freeze
the cineloop.
2.
Use the
Trackball
to scroll through the cineloop frame by
frame.
Or
Use the Speed/Frame softkey button.
Содержание Vivid i
Страница 110: ...Scanning Modes 100 Vivid i User s Manual 2378958 100 Rev 02...
Страница 111: ...Purposely left empty Vivid i User s Manual 101 2378958 100 Rev 02 Chapter 4 Purposely left empty For Future purposes...
Страница 112: ...Purposely left empty 102 Vivid i User s Manual 2378958 100 Rev 02 To Purchase Visit Avobus com or call 1 800 674 3655...
Страница 161: ...Purposely Left Empty Vivid i User s Manual 151 2378958 100 Rev 02 Chapter 7 Purposely Left Empty For Future purposes...
Страница 162: ...Purposely Left Empty 152 Vivid i User s Manual 2378958 100 Rev 02...
Страница 288: ...Probes 278 Vivid i User s Manual 2378958 100 Rev 02...
Страница 350: ...User maintenance 340 Vivid i User s Manual 2378958 100 Rev 02...
Страница 376: ...Safety 366 Vivid i User s Manual 2378958 100 Rev 02 To Purchase Visit Avobus com or call 1 800 674 3655...
Страница 387: ......
Страница 388: ...To Purchase Visit Avobus com or call 1 800 674 3655...