Measurement and Analysis
132
Vivid i User’s Manual
2378958-100 Rev. 02
Acceleration, Acceleration time (AT)
1.
Select
Accel
or
AT
.
2.
Position the caliper at the start point and press
SET
to
anchor the caliper.
3.
Position the second caliper at the end point and press
SET
to anchor the caliper and complete the measurement.
The acceleration and/or the acceleration time is displayed
in the
Measurement result table
.
Heart rate
Heart rate is calculated by selecting two identical points over
four heart cycles.
1.
Select
HR
.
2.
Position the caliper at a recognizable point in the first cycle
and press
SET
to anchor the caliper.
3.
Position the second caliper at the identical point in the
fourth cycle and press
SET
to anchor the caliper and
complete the measurement.
The Heart rate is displayed in the
Measurement result
table
.
Peak systole (PS), End diastole (ED) and Mid diastole (MD)
1.
Select
PS
,
ED
or
MD
.
2.
Position the caliper at the corresponding measurement
point and press
SET
to complete the measurement.
The selected measurement is displayed in the
Measurement result table
.
Pulsatility index (PI)
With Auto trace on
1.
Select
PI
.
2.
Position the caliper at the beginning of the wave form and
press
SET
to anchor the caliper.
3.
Position the second caliper at end diastole and press
SET
.
A trace is displayed between the two calipers and PS, MD,
ED, TAMAX and PI are displayed in the
Measurement
result table
.
With Manual trace on
1.
Select
PI
.
2.
Position the caliper at the beginning of the wave form and
press
SET
to anchor the caliper.
To Purchase, Visit
Содержание Vivid i
Страница 110: ...Scanning Modes 100 Vivid i User s Manual 2378958 100 Rev 02...
Страница 111: ...Purposely left empty Vivid i User s Manual 101 2378958 100 Rev 02 Chapter 4 Purposely left empty For Future purposes...
Страница 112: ...Purposely left empty 102 Vivid i User s Manual 2378958 100 Rev 02 To Purchase Visit Avobus com or call 1 800 674 3655...
Страница 161: ...Purposely Left Empty Vivid i User s Manual 151 2378958 100 Rev 02 Chapter 7 Purposely Left Empty For Future purposes...
Страница 162: ...Purposely Left Empty 152 Vivid i User s Manual 2378958 100 Rev 02...
Страница 288: ...Probes 278 Vivid i User s Manual 2378958 100 Rev 02...
Страница 350: ...User maintenance 340 Vivid i User s Manual 2378958 100 Rev 02...
Страница 376: ...Safety 366 Vivid i User s Manual 2378958 100 Rev 02 To Purchase Visit Avobus com or call 1 800 674 3655...
Страница 387: ......
Страница 388: ...To Purchase Visit Avobus com or call 1 800 674 3655...