Measurement and Analysis
122
Vivid i User’s Manual
2378958-100 Rev. 02
4.
Select
Ao/LA
in the
Measurement Menu
.
5.
Trackball the cursor along the time axis to the required point
to start measurement of Aorta root diameter.
See the
Status bar
to get the next step
to perform.
6.
Press
SET
.
The starting point for the measurement is anchored.
7.
Trackball to the end point of the measurement.
The current value is
updated while mov-
ing the cursor.
8.
Press
SET
.
The measurement end point is anchored and the value is
displayed in the
Measurement result table
.
A new free-moving cursor is displayed on the image, ready
for the next measurement.
9.
Repeat steps 5, through 8 to measure Left Atrium.
The LA value is displayed in the
Measurement result table
.
The Ao/LA ratio is displayed in the
Measurement result
table
.
LV study
The LV study consists of measurements in fixed-time mode in
both systole and diastole of:
•
Interventricular septum thickness (IVS)
•
Left ventricular internal dimension (LVID)
•
Left ventricular posterior wall thickness (LVPW)
The following parameters are also calculated:
•
EDV (End diastole volume)
•
ESV (End systole volume)
•
SV (Stroke volume)
•
EF (Ejection Fraction)
To perform LV study
1.
Generate the M-Mode image.
2.
Press
FREEZE
to stop the cineloop.
3.
Press
MEASURE
on the Control Panel.
4.
Select
LV study
in the
Measurement Menu
.
5.
Trackball the cursor along the time axis to the required point
to start measurement of IVSd.
6.
Press
SET
. The starting point for the measurement is
anchored.
7.
Trackball to the end point of the measurement.
Содержание Vivid i
Страница 110: ...Scanning Modes 100 Vivid i User s Manual 2378958 100 Rev 02...
Страница 111: ...Purposely left empty Vivid i User s Manual 101 2378958 100 Rev 02 Chapter 4 Purposely left empty For Future purposes...
Страница 112: ...Purposely left empty 102 Vivid i User s Manual 2378958 100 Rev 02 To Purchase Visit Avobus com or call 1 800 674 3655...
Страница 161: ...Purposely Left Empty Vivid i User s Manual 151 2378958 100 Rev 02 Chapter 7 Purposely Left Empty For Future purposes...
Страница 162: ...Purposely Left Empty 152 Vivid i User s Manual 2378958 100 Rev 02...
Страница 288: ...Probes 278 Vivid i User s Manual 2378958 100 Rev 02...
Страница 350: ...User maintenance 340 Vivid i User s Manual 2378958 100 Rev 02...
Страница 376: ...Safety 366 Vivid i User s Manual 2378958 100 Rev 02 To Purchase Visit Avobus com or call 1 800 674 3655...
Страница 387: ......
Страница 388: ...To Purchase Visit Avobus com or call 1 800 674 3655...