Measurement and Analysis
134
Vivid i User’s Manual
2378958-100 Rev. 02
The trace is displayed and the corresponding
measurements are displayed in the
Measurement result
table
.
Measurement package configuration
A list of all cardiac calculations with needed measurements and
location in the Measurement package can be found in the
Reference manual.
There are many more measurements and parameters in the
measurement package than shown in the default
Measurement
menu
. Use the configuration system to set up the
measurements that should be available in the
Measurement
menu
and which parameters should be calculated.
The following example based on calculation of AV CO (Cardiac
Output by Aortic Flow) describes how to configure the
measurement package so that necessary measurements and
the resulting calculations are displayed on screen.
Measurement package configuration -
example
Calculation of Cardiac Output by Aortic Flow requires
measurement of:
•
AV diameter located in the folder
Dimension
(2D mode)
•
AV VTI located in the folder
Aortic
(Doppler AV Trace).
• Heart
rate
If a calculated parameter (e.g. AV CO in AV Trace
measurement) requires another parameter to be calculated
(e.g. AV Diam) the user must first measure the required
parameter (e.g. AV Diam) before the dependent parameter (e.g
AV CO in AV Trace) gets calculated.
Configuration of the Measurement menu
If the AV diameter measurement is not present in the folder
Dimension
in the
Measurement menu
, follow the following
procedure:
1.
Press
CONFIG
and select the category
Measure
.
The
Measurement menu sheet
is displayed (see
Содержание Vivid i
Страница 110: ...Scanning Modes 100 Vivid i User s Manual 2378958 100 Rev 02...
Страница 111: ...Purposely left empty Vivid i User s Manual 101 2378958 100 Rev 02 Chapter 4 Purposely left empty For Future purposes...
Страница 112: ...Purposely left empty 102 Vivid i User s Manual 2378958 100 Rev 02 To Purchase Visit Avobus com or call 1 800 674 3655...
Страница 161: ...Purposely Left Empty Vivid i User s Manual 151 2378958 100 Rev 02 Chapter 7 Purposely Left Empty For Future purposes...
Страница 162: ...Purposely Left Empty 152 Vivid i User s Manual 2378958 100 Rev 02...
Страница 288: ...Probes 278 Vivid i User s Manual 2378958 100 Rev 02...
Страница 350: ...User maintenance 340 Vivid i User s Manual 2378958 100 Rev 02...
Страница 376: ...Safety 366 Vivid i User s Manual 2378958 100 Rev 02 To Purchase Visit Avobus com or call 1 800 674 3655...
Страница 387: ......
Страница 388: ...To Purchase Visit Avobus com or call 1 800 674 3655...