Measurement and Analysis
Vivid i User’s Manual
123
2378958-100 Rev. 02
8.
Press
SET
. The IVSd measurement end point is anchored
and the value is displayed in the
Measurement result table
.
The end point of the IVSd is also the start point for the LVIDd.
1.
Trackball to the end point of the LVIDd measurement.
2.
Press
SET
. The LVIDd measurement end point is anchored
and the value is displayed in the
Measurement result table
.
The end point of the LVIDd is also the start point for the
LVPWd.
1.
Trackball to the end point of the LVPWd measurement.
2.
Press
SET
. The LVPWd measurement end point is
anchored and the value is displayed in the
Measurement
result table
.
3.
Repeat steps 5, through 2 to measure IVS, LVID and LVPW
in systole.
RV study
The RV study consists of measurement in fixed-time mode of
Right ventricular internal dimension (RVID) in both diastole and
systole.
To perform RV study
1.
Generate the M-Mode image.
2.
Press
FREEZE
to stop the cineloop.
3.
Press
MEASURE
on the Control Panel.
4.
Select
RV study
in the
Measurement Menu
.
5.
Trackball the cursor along the time axis to the required point
to start measurement of RVIDd.
6.
Press
SET
. The starting point for the measurement is
anchored.
The current value is
updated while mov-
ing the cursor.
7.
Trackball to the end point of the measurement.
8.
Press
SET
. The measurement end point is anchored and
the RVIDs measurement value is displayed in the
Measurement result table
.
A new free-moving cursor is displayed on the image, ready
for the next measurement.
9.
Repeat steps 5, through 8 to measure RVIDs.
The RVIDs value is displayed in the
Measurement result
table
.
To Purchase, Visit
Содержание Vivid i
Страница 110: ...Scanning Modes 100 Vivid i User s Manual 2378958 100 Rev 02...
Страница 111: ...Purposely left empty Vivid i User s Manual 101 2378958 100 Rev 02 Chapter 4 Purposely left empty For Future purposes...
Страница 112: ...Purposely left empty 102 Vivid i User s Manual 2378958 100 Rev 02 To Purchase Visit Avobus com or call 1 800 674 3655...
Страница 161: ...Purposely Left Empty Vivid i User s Manual 151 2378958 100 Rev 02 Chapter 7 Purposely Left Empty For Future purposes...
Страница 162: ...Purposely Left Empty 152 Vivid i User s Manual 2378958 100 Rev 02...
Страница 288: ...Probes 278 Vivid i User s Manual 2378958 100 Rev 02...
Страница 350: ...User maintenance 340 Vivid i User s Manual 2378958 100 Rev 02...
Страница 376: ...Safety 366 Vivid i User s Manual 2378958 100 Rev 02 To Purchase Visit Avobus com or call 1 800 674 3655...
Страница 387: ......
Страница 388: ...To Purchase Visit Avobus com or call 1 800 674 3655...