D
IRECTION
5771498-100, R
EVISION
5
V
ENUE
™ S
ERVICE
M
ANUAL
Chapter 5 - Venue™Components and Function (Theory)
5-29
P R E L I M I N A R Y
Section 5-11
Cooling System
5-11-1
General Information
The Venue™ system has 3 fans for system cooling:
•
Back End Fan - located on the Back End module.
•
T-CFE Fan - located on the top of the Front End module.
•
MPB Blower (Fan) - located on the on the rear side of the system behind the MPB module.
All fans are controlled and have variable speed.
The cooling requirement for the Venue™ ultrasound scanner with monitor and on board peripherals, is
up to 2000 BTU/h. This figure does not include cooling needed for lights, people, or other equipment in
the room.
NOTE:
Each person in the room places an additional 300 BTU/h demand on the cooling system.
Содержание Venue R1
Страница 477: ......