Chapter 8 - Replacement Procedures
8-173
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
8-8-3
ECG Installation Procedure for Systems with No Printer Installed
8-8-3-1
Tools
Appropriate Phillips screwdriver
8-8-3-2
Time Required
15 min
8-8-3-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
.
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-8-3-4
ECG Installation
1.) Remove the following covers: Lower Front eTower, Upper Front eTower, and MPB Front Metal
Door, Printer Insert Cover.
2.) Replace the printer insert cover with the one provided in the ECG installation kit.
FRU Part #
Refer to
.
•
Lower Front eTower Cover Removal Procedure
•
Upper eTower Front Cover Removal Procedure
•
MPB Front Metal Door Removal Procedure
•
Printer Insert Cover Removal Procedure
Figure 8-185 Replacing Printer Insert Cover