8-50
Section 8-4 - System Modules- Replacement Procedures
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
8-4-9
BEP PCB Replacement Procedure
8-4-9-1
Tools
•
Appropriate Phillips screwdriver
8-4-9-2
Time Required
15 min
8-4-9-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
.
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-4-9-4
BEP PCB Removal Procedure
1.) Remove all accessories.
2.) Remove the following covers: Lower Front eTower, Left side eTower, and Right Side eTower,
3.) Perform BE module removal procedure.
4.) Remove omega bracket: release 5 screws connecting the omega bracket:
5.) Remove the supporting bracket using an appropriate Phillips screwdriver:
•
Accessories - Replacement Procedures
•
RS Probe Cover Removal Procedure
•
Lower Front eTower Cover Removal Procedure
•
Left Side eTower Cover Removal Procedure
•
Right Side eTower Cover Removal Procedure
•
MPB Front Metal Door Removal Procedure
•
Figure 8-36 BE Omega Bracket Removal
Содержание Venue R1