Error and System Message
74
Marquette Responder® 3000
227 490 02-C
17 Error and System Messages
Message
Effect
Explanation
Remedy
D-RAM Error
device defect, device
cannot be used
notify service
S-RAM Error
device defect, device
cannot be used
notify service
FLASH Error
device defect, device
cannot be used
notify service
EEPROM Error
restricted use for
emergencies only,
defibrillation possible
defective user setup (e.g.
user setup cannot be
saved)
press one of the function keys,
notify service
Charge Energy Error
charged energy differs
from selected energy, use
device only in emergen-
cies
error occurred last time
the device was used
(before switched off) and
was saved
press one of the function keys,
trigger test discharge, switch
device off and on again; if error
message recurs, notify service
Time Base Error
restricted use for
emergencies only,
defibrillation possible
all time-related data (e.g
heart rate) may be
erroneous
press one of the function keys,
adjust time; if error message
recurs, notify service
Battery
battery almost depleted
charge battery
HR, etC
alarm, violation of alarm
limit
Electrode
high ECG or defibrillation
electrode impedance
reapply electrode
Check selector
energy selector is not
indexing properly;
defibrillation possible
set energy selector to the exact
position; if error message recurs,
notify service
Energy high
internal spoons are
connected and selected
energy is above 50 J;
charged energy higher
than selected energy;
defibrillation possible
select a lower value (internal
spoons) or trigger test discharge;
if message recurs, notify service
Energy low
charged energy below
selected energy;
defibrillation possible
(e.g. battery depleted)
trigger test discharge; if message
recurs, notify service
check battery
Device Behavior
Explanation
Remedy
Device not functioning
properly
software problem