Manual Defibrillation / Non-synchronized Defibrillation
20
Marquette Responder® 3000
227 490 02-C
SHOCK
3
STERNUM
SHOCK
3
CHARGE
2
APEX
a
b
Figure 4-4.
Buttons to initiate defibrillator
charging (a) and to trigger the shock
(a+b)
Warning
Risk of Skin Burns / Equipment Damage — Do
not apply the paddles over
-
sternum or clavicle,
-
nipples
-
implanted pacemaker or defibrillator devices.
Note
−
If you do not trigger the shock within 30 sec-
onds of charging, the energy will automatically
be discharged internally. You will then have to
recharge the defibrillator.
−
When the defibrillator is charged, you can
increase and decrease the energy level to any
value with the energy selector (without pressing
the "Charge" button again).
For an internal discharge of the stored energy,
set the energy selector to
or to "Off".
•
Initiate energy storage with the button on the
APEX paddle (a).
When the selected energy is stored,
−
the device emits an audio signal
−
the message "Energy available" appears,
−
the available energy is displayed (if the
available energy drops below a given level, the
defibrillator recharges automatically).
•
Now trigger the shock within 30 seconds. To do
so, simultaneously press the buttons a and b on
the paddles.
After the shock
−
the audio signal stops and the delivered energy
is displayed for approx. 6 seconds (in place of
the available energy),
−
the recorder prints a 14-second ECG strip (4
seconds before the shock, 5 seconds blanked,
10 seconds after the shock) (configurable); the
blanked period of time is indicated by a vertical
line on the recording;
−
the shock delivery is annotated on the stored
ECG (also refer to chapter 11 "Memories of the
Marquette Responder® 3000").
If the defibrillator cannot store the selected energy
so that selected and stored energy values differ, a
warning will be displayed. The defibrillation pulse
can be triggered all the same.
In this situation we recommend to check the
batteries first. If the batteries are intact, have the
defibrillator immediately repaired.