Manual Defibrillation / Non-Synchronized Defibrillation
227 490 02-C
Marquette Responder® 3000
25
3
2
Analyse
Sync
Figure 4-12. Buttons to initiate defibrillator
charging and to trigger the shock
Defibrillating the Patient
With internal electrodes, it is not possible to select
a value above 50 Joules, because higher energies
may damage the myocardium. When you set the
energy selector to a value above 50 Joules, you
will be alerted by a message and defibrillator
charging will not proceed.
•
Defibrillate the patient as described for
defibrillation with paddles (page 19).
Be sure to charge the defibrillator with
and
to deliver the shock by simultaneously pressing
buttons
and
(Figure 4-12).