Cleaning, Maintenance
76
Marquette Responder® 3000
227 490 02-C
Figure 18-1. Inserting the spoon electrodes
Note
Having loosened the counter nut 2, you can easily
alter the position of the contact paddle.
Figure 18-2. Counter electrode for internal
defibrillation
Inserting the Spoon Electrode
•
Screw the counter nut 2 (Figure 18-1) onto the
electrode as far as it will go.
•
Screw the contact paddle 1 into the handle as
far as it will go, then bring it into the appropri-
ate position.
•
Now fix the contact paddle by screwing the
counter nut 2 tight against the handle 3.
External Counter Electrode for Internal
Defibrillation
•
Disconnect the electrode from its lead before
cleaning or sterilizing it (Figure 18-2).
•
Clean the electrode by rubbing it down with a
cloth moistened with soap water. Use a disin-
fectant for disinfection. Do not immerse the
electrode in the liquid.
•
Low-temperature plasma sterilization is the
recommended sterilization method. Alternative
methods are ETO sterilization or ionizing
radiation. (Please note: Frequent ETO steriliza-
tion reduces the life expectancy of the plastic
material!) Do not autoclave the electrodes.