Pulse Oximetry
52
Marquette Responder® 3000
227 490 02-C
ECG
I
II
III
15.07.1999 09:05:00
160 / 40
bpm
semiautom.
89
SpO2
---
90
%
97
ECG
SpO2
etCO2
QRSPulse
Tone
OFF
Next
Menu
Figure 9-2.
SpO
2
parameter window
OFF
Time
8 s
Tone
OFF
C-LOCK
Integ.
QRSPulse
Previous
Menu
Source
ECG
HR
Figure 9-3.
SpO
2
menu
Note
Disable the C-Lock ECG synchronization feature
when monitoring pacemaker patients.
Warning
Delayed Alarm — When monitoring the pulse rate
derived from the SpO
2
signal instead of the heart
rate, select an integration time of 4 or 8 seconds;
do not select 12 seconds because of the slower
display update.
Measuring and Monitoring Oxygen
Saturation
•
Apply the probe as described in the user
instructions supplied with the probe and con-
nect it to the defibrillator (blue connector).
The SpO
2
measurement begins as soon as the
sensor is connected. During the short self-test, the
parameter window is blue. As soon as the probe
supplies a valid signal, the color changes to green
(Figure 9-2).
At the factory, only a low limit of 90% is adjusted,
no high limit. When the parameter reading drops
below this value, the defibrillator sounds an alarm
and the window changes from green to red.
Enabling/Disabling C-Lock ECG Synchroniza-
tion
With the factory default settings unchanged, the C-
Lock ECG synchronization is disabled when the
device is first turned on. If the ECG signal is
acquired, enable the feature.
•
Press F2
SpO2 etCO2
.
•
Press F3
SpO2
to display the SpO
2
menu
(Figure 9-3) (displays only when SpO
2
probe is
connected).
•
Press F1
C-LOCK
to enable the C-Lock feature
(or press the key again to disable it).
Selecting the Integration Time
The integration time is the time over which the
measured values are averaged. You can select an
integration time of 4, 8 or 12 seconds. The default
setting is 8 seconds. The integration time of 12
seconds should only be selected in exceptional
cases.
•
Display the SpO
2
menu as described above
(Figure 9-3).
•
Select the integration time with F2
Integ. Time
.