Battery Power Operation
227 490 02-C
Marquette Responder® 3000
67
Figure 14-1. Battery charging indicators
Warning
-
Operational Readiness — Defibrillators are
emergency medical devices designed to save
and preserve lives. They must be ready for use
at all times. This applies also to mains-
independent operation. Ensure that the device
battery is always fully charged.
-
Do not recharge NiCd batteries in direct
sunlight, on sources of heat or at extremely
low ambient temperatures (minimum tem-
perature 5 °C/41 °F). The ambient tempera-
ture should not exceed 40 °C/104 °F, as this
could have adverse effects on the battery's
service life.
14 Battery Power Operation
Battery Charging (ambulance power supply,
power line)
The batteries recharge when you
−
place the defibrillator in the vehicle mounting
unit / wall mount unit or
−
connect the defibrillator to the power line (units
with AC power adapter).
The two indicators (Figure 14-1) light up when the
batteries are charging. Charging will take approx.
4 hours per battery (the batteries are recharged one
after the other). The defibrillator can be operated
in the vehicle mounting system or on mains power,
even when the batteries are discharged. Also
observe the information given in chapter 16
"Operation in the Vehicle Mounting Unit".
Charging the Batteries with the Separate
Charging Unit ASU 3000
In addition to charging the batteries, the charging
unit also maintains them (see section "Information
on Battery Power Operation" below). Please note
the instructions given in the charging unit
operator's manual.