12-Lead ECG Analysis Program (12SL
TM
)
48
Marquette Responder® 3000
227 490 02-C
Start
ECG
I
15.07.1998 09:05:00
160 / 40
bpm
Recall
Previous
Menu
semiautom.
62
Figure 8-1.
12SL analysis menu
Recall
ECG
I
20.07.1998 14:35:37
160 / 40
bpm
Recall
20.07.1998 14:35:37
NORMAL SINUS RHYTHM
NORMAL ECG
Recall
Previous
Menu
semiautom.
62
Figure 8-2.
12SL interpretation
12SL Measurement
For measurement and interpretation 12 seconds of
ECG data are saved. After the analysis, the
interpretation is presented on the monitor screen.
You can also print the ECG including the
measurement results and the interpretation. The
printed segment, however, is only 3,2 seconds.
If 12SL analysis was already performed with the
defibrillator when you activate the program, the
previous interpretation will appear on the display.
In addition, the full-disclosure ECG of channel 1 is
stored in the ECG memory.
•
Press F1
ECG
.
•
Display the 12SL menu with F4
12SL
(Figure
8-1).
•
Press F1
Start
to start the program.
After acquisition, the ECG data is analyzed and
stored. Afterwards, the interpretation is displayed
on the screen (Figure 8-2).
Press F2
to print the ECG, including
measurement results and interpretation.
Press F4
Recall
to recall the 3 most recent 12SL
records. The corresponding acquisition time is
shown in the softkey.