Cleaning, Maintenance
227 490 02-C
Marquette Responder® 3000
77
Maintenance
Checks before each use
•
Before each use, visually inspect the device, the
leads and electrodes for signs of mechanical
damage.
•
Also check the device performance as outlined
in chapter 3 "Putting the Device Into Opera-
tion", and trigger a test discharge (chapter 15).
If you detect damages or impaired functions which
may result in a hazard to the patient or the
operator, the device must be repaired before it can
be used again.
Checks at regular intervals
The Marquette Responder® 3000 is an emergency
device and must always be ready for use. These
checks must be carried out once a month:
•
visual inspection of the device and the
accessories
•
performance check as outlined in chapter 3
"Putting the Device Into Operation"
Technical Inspections
For safety, the devices require regular mainte-
nance. To ensure functional and operational safety
of the Marquette Responder® 3000, Technical
Inspections should be carried out on an annual
basis.
These checks should be performed by persons with
adequate training and experience.
The checks can be carried out by Marquette
Hellige within the framework of a service contract.
The nature and scope of these checks are ex-
plained in the corresponding sections of the
Service Manual.
Battery Replacement
NiCd batteries have a limited service life as their
storage capacity deteriorates with age. For this
reason, the batteries must be replaced every 3
years.
The device does not require any other mainte-
nance.
Disposal at the End of Its Service Life
Note
At the end of their service life, the device described
in this manual and its accessories must be dis-
posed of in compliance with the applicable local
waste control regulations. If you have questions
regarding the disposal of the product or of the
accessories, please contact Marquette Hellige
GmbH or its representatives.