Putting the Device Into Operation and Performance Check
12
Marquette Responder® 3000
227 490 02-C
Figure 3-1.
Defibrillator with AC power adapter
Note
The Marquette Responder® 3000 is switched on
and off with the energy selector. Once you are
familiar with the device operating routines, this
control lets you turn on the device and select the
required defibrillation energy in one action. No
shock can be delivered in the
position of the
energy selector.
1
2
5
7
10
20 30
50
100
150
200
300
360
Autoseq
Figure 3-2.
Turning the defibrillator on
Power Supply
The defibrillator can be powered
−
from the power line (requires AC power
adapter, P/N 205 108 01, Figure 3-1),
−
from the ambulance power supply system
(requires vehicle mounting unit,
P/N 202 317 01),
−
from the wall mount unit (P/N 202 317 03)
−
from 1 or 2 rechargeable batteries (mains-
independent).
Batteries are recharged by one of the following
methods:
−
in the defibrillator, when the defibrillator is
connected to the power line or to the ambulance
power supply system, or
−
by means of the separate charging unit ASU
3000 (P/N 701 279 01).
If you prefer to operate the device mains-
independent, ensure that the batteries are charged
(chapter 14 "Battery Power Operation").
Please refer to chapter 16 for information on
operating the device in the vehicle mounting unit
and on installing the AC power adapter.
Turning the Defibrillator On
•
Connect the device to the power supply.
•
Switch on the device by turning the energy
selector to
(defibrillation shocks cannot be
delivered in this position).
The test screen appears and the device emits a
short audio signal.
On the test screen you can see the software version
and a message referring to the self-test. The three
color blocks in red, green and blue are displayed to
verify the correct representation of the colors.
After the self-test the standard screen appears
(Figure 3-3).